HPLC — High-Performance Liquid Chromatography
A machine that separates a mixture of compounds by pumping it through a packed column under high pressure. Each compound sticks to the column material differently, so each one exits at a different time — that time is its identity.
The mobile phase (solvent) is driven by a high-pressure pump through a column packed with stationary phase particles (typically silica-based, C18 reversed-phase). Separation is governed by differential partitioning between the two phases. Column efficiency is described by the van Deemter equation:
H = A + B/u + Cu
where H = height equivalent of a theoretical plate, u = linear velocity. Minimizing H maximizes theoretical plate count (N) and resolution (Rs).
| Parameter | Typical Range |
| Particle size | 3–10 µm |
| Max pressure | ~400 bar / 6,000 psi |
| Flow rate | 0.5–2.0 mL/min |
| Column length | 50–250 mm |
UPLC — Ultra-Performance Liquid Chromatography
The same principle as HPLC, but with much smaller column particles — which means faster, sharper separations, but requires much higher pressure.
Reducing particle size (dp < 2 µm) dramatically lowers H per the van Deemter equation, increasing N per unit column length. The trade-off: backpressure increases as ~1/dp², requiring pump and hardware rated to 1,200 bar / 17,000 psi.
| Parameter | HPLC | UPLC |
| Particle size | 3–10 µm | < 2 µm |
| Max pressure | ~400 bar | ~1,200 bar |
| Run time | 15–60 min | 2–10 min |
| Solvent use | High | ~80% less |
| Resolution | Good | Superior |
UV-Vis Detector
Shines a single wavelength of UV light through the sample stream. If the compound absorbs that light, a signal is generated. The stronger the signal, the higher the concentration.
Operates on the Beer-Lambert Law: A = ε · c · l, where A = absorbance, ε = molar absorptivity, c = concentration, l = path length. Response is linear with concentration within a defined range. Detection requires the analyte to possess a chromophore.
| Strength | Limitation |
| Simple, robust, inexpensive | Single wavelength only |
| Linear response | Requires chromophore |
| Widely validated (pharmacopoeial) | No spectral confirmation |
DAD / PDA — Are They Different?
Both measure absorbance across the full UV-Vis spectrum simultaneously. The only difference is the name: DAD (Agilent/Shimadzu) = PDA (Waters/Thermo). Same hardware, different brand terminology.
An array of hundreds of photodiodes captures the full spectrum (190–800 nm) at every time point, generating a 3D data matrix (time × wavelength × absorbance). Enables: (1) peak purity analysis; (2) spectral library matching.
| Feature | UV-Vis | DAD / PDA |
| Wavelengths | 1 (fixed) | Full spectrum simultaneously |
| Peak purity check | ❌ | ✅ |
| Library matching | ❌ | ✅ |
| 3D chromatogram | ❌ | ✅ |
RI Detector — Refractive Index
Measures how much the dissolved compound bends light compared to pure solvent. No chromophore needed — but it can't be used when the solvent composition is changing (gradient elution).
A differential refractometer compares the refractive index (n) of the reference cell against the sample cell. Response: Δn = n_sample − n_reference, proportional to analyte concentration. Incompatible with gradient methods due to baseline drift. Sensitivity is inherently low (LOD ~ µg range).
| Strength | Limitation |
| No chromophore needed | Incompatible with gradient |
| Near-universal (non-volatile compounds) | Low sensitivity |
| Simple quantitation | Temperature-sensitive baseline |
FLD — Fluorescence Detector
Some compounds absorb light at one wavelength and re-emit at a longer wavelength. The detector measures only the emitted light — extremely selective and 100–1,000× more sensitive than UV-Vis.
The analyte absorbs photons at λex, electrons promoted to S1 state, relaxation emits at λem > λex (Stokes shift). Only compounds that both absorb AND emit respond — matrix background dramatically suppressed. When native fluorescence is absent, derivatization introduces a fluorophore tag.
| Strength | Limitation |
| Highest sensitivity of all optical detectors | Not all compounds fluoresce |
| Excellent selectivity | May require derivatization |
| Compatible with gradient | More complex optimization |
ELSD — Evaporative Light Scattering Detector
The solvent is evaporated away, leaving compound particles. A laser scatters off those particles — more particles = stronger signal. Works for any non-volatile compound regardless of UV absorption.
Three stages: (1) Nebulization — eluent → aerosol by N₂ gas; (2) Evaporation — mobile phase removed in heated drift tube; (3) Detection — particles scatter laser/LED beam. Response follows power law: Response = a · mb (non-linear). Fully gradient-compatible unlike RI.
| Strength | Limitation |
| No chromophore required | Non-linear response |
| Gradient-compatible | Analyte must be non-volatile |
| Near-universal | Lower sensitivity than FLD |
Full Detector Comparison
| Detector | Chromophore? | Gradient? | Sensitivity | Best Use |
| UV-Vis | ✅ Yes | ✅ Yes | Moderate | Drugs, aromatics |
| DAD/PDA | ✅ Yes | ✅ Yes | Moderate | Purity, libraries |
| RI | ❌ No | ❌ No | Low | Sugars, polymers |
| FLD | ❌ No* | ✅ Yes | Very high | Trace contaminants |
| ELSD | ❌ No | ✅ Yes | Moderate | Lipids, natural products |
HPLC — كروماتوغرافيا السائل عالية الأداء
جهاز يفصل مكونات خليط بضخّها عبر عمود مضغوط. كل مركب يتفاعل مع مادة العمود بشكل مختلف فيخرج في وقت مختلف — هذا الوقت هو هويته.
تدفع المضخة الطور المتحرك عبر عمود محشو بجسيمات الطور الثابت (غالباً سيليكا C18). الفصل يعتمد على الانقسام التفاضلي بين الطورين. كفاءة العمود تُحكَم بمعادلة van Deemter:
H = A + B/u + Cu
| المعامل | النطاق النموذجي |
| حجم الجسيمات | 3–10 µm |
| أقصى ضغط | ~400 bar |
| معدل التدفق | 0.5–2.0 mL/min |
| طول العمود | 50–250 mm |
UPLC — كروماتوغرافيا السائل فائقة الأداء
نفس مبدأ الـ HPLC، لكن بجسيمات عمود أصغر بكثير — فصل أسرع وأحدّ، لكن يتطلب ضغطاً أعلى بكثير.
| المعامل | HPLC | UPLC |
| حجم الجسيمات | 3–10 µm | < 2 µm |
| أقصى ضغط | ~400 bar | ~1,200 bar |
| وقت التحليل | 15–60 دقيقة | 2–10 دقائق |
| استهلاك المذيب | مرتفع | أقل بـ ~80% |
| جودة الفصل | جيدة | ممتازة |
كاشف UV-Vis
يسلّط ضوءاً بطول موجي محدد على تيار العينة. إذا امتص المركب هذا الضوء، تُولَّد إشارة. قوة الإشارة = التركيز.
يعمل وفق قانون Beer-Lambert: A = ε · c · l. يشترط وجود chromophore في الجزيء.
| نقاط القوة | نقاط الضعف |
| بسيط وموثوق وغير مكلف | طول موجي واحد فقط |
| استجابة خطية | يشترط وجود Chromophore |
| معتمد فارماكوبيائياً | لا يوفر تأكيداً طيفياً |
DAD مقابل PDA
كلاهما يقيس الامتصاصية عبر الطيف الكامل في آنٍ واحد. الفرق الوحيد هو الاسم: DAD عند Agilent وShimadzu، وPDA عند Waters وThermo.
| الخاصية | UV-Vis | DAD / PDA |
| الأطوال الموجية | 1 (ثابت) | الطيف الكامل |
| فحص نقاء الـ Peak | ❌ | ✅ |
| مطابقة المكتبة | ❌ | ✅ |
| كروماتوغرام ثلاثي الأبعاد | ❌ | ✅ |
كاشف RI — معامل الانكسار
يقيس كيف يُغيّر المركب المذاب انكسار الضوء مقارنةً بالمذيب النقي. لا يحتاج Chromophore — لكنه لا يعمل مع Gradient.
| نقاط القوة | نقاط الضعف |
| لا يحتاج Chromophore | غير متوافق مع Gradient |
| شبه شامل | حساسية منخفضة |
| قياس كمي بسيط | Baseline حساس للحرارة |
كاشف FLD — الفلورة
بعض المركبات تمتص ضوءاً ثم تُعيد إصداره عند طول موجي أعلى. الكاشف يقيس الضوء المُعاد إصداره فقط — انتقائي للغاية وأعلى حساسية بـ 100–1,000 مرة من UV-Vis.
| نقاط القوة | نقاط الضعف |
| أعلى حساسية | ليس كل المركبات فلورية |
| انتقائية ممتازة | قد يحتاج Derivatization |
| متوافق مع Gradient | تحسين أكثر تعقيداً |
كاشف ELSD — التشتت الضوئي بالتبخير
يُبخَّر المذيب بالكامل، تبقى جزيئات المركب الصلبة. شعاع ليزر يتشتت عليها — يعمل لأي مركب غير متطاير.
| نقاط القوة | نقاط الضعف |
| لا يحتاج Chromophore | استجابة غير خطية |
| متوافق مع Gradient | يشترط مركب غير متطاير |
| شبه شامل | حساسية أقل من FLD |
جدول المقارنة الشامل
| الكاشف | Chromophore؟ | Gradient؟ | الحساسية | الأنسب لـ |
| UV-Vis | ✅ نعم | ✅ نعم | متوسطة | أدوية، مركبات عطرية |
| DAD/PDA | ✅ نعم | ✅ نعم | متوسطة | تأكيد النقاء |
| RI | ❌ لا | ❌ لا | منخفضة | سكريات، بوليمرات |
| FLD | ❌ لا | ✅ نعم | عالية جداً | ملوثات، أفلاتوكسين |
| ELSD | ❌ لا | ✅ نعم | متوسطة | دهون، منتجات طبيعية |